SFDA Reliance Pathway for Medical Devices in Saudi Arabia 2026

SFDA's MDS-G030 reliance pathway (Sep 2026) lets devices approved by TGA, Health Canada, EU, PMDA, MHRA or FDA get a Saudi MDMA in 35-60 working days.

Last reviewed: 8 min read By the Mutamad regulatory team

On 9 September 2026, the Saudi Food and Drug Authority (SFDA) published MDS-G030 v1.0, a new reliance pathway for medical device marketing authorization (MDMA). If an identical device is already authorized by TGA, Health Canada, an EU/EEA national competent authority, PMDA, MHRA or the US FDA, SFDA can rely on that decision instead of performing a full technical review, with a stated review time of 35–60 working days. Self-declared, 510(k)-type, custom-made and safety-action devices are excluded, and Saudi-specific requirements still apply.

This guide explains what the pathway is, who qualifies, what is excluded and how it compares with the standard MDMA route.

What “reliance” means at SFDA

In regulatory terms, reliance means one regulator takes account of another trusted regulator’s assessment when it makes its own decision. SFDA still issues the MDMA and remains responsible for it. What changes is the depth of review: SFDA uses the reference authority’s authorization as evidence that the device is safe and performs as intended, rather than re-reviewing the full technical file.

This matters for manufacturers with devices already on the market in major jurisdictions. Before MDS-G030, those manufacturers normally went through the standard route: a full technical documentation review against MDS-REQ 1. The official guidance is on the SFDA website.

Reliance is not automatic recognition. You still need an MDMA, a Saudi Authorized Representative and Saudi-compliant labelling.

Reference authorities

MDS-G030 names six reference regulators:

Reference authority Jurisdiction
TGA (Therapeutic Goods Administration) Australia
Health Canada Canada
National competent authorities EU / EEA
PMDA (Pharmaceuticals and Medical Devices Agency) Japan
MHRA (Medicines and Healthcare products Regulatory Agency) United Kingdom
US FDA (Food and Drug Administration) United States

The UK (MHRA) is on the list in its own right, separately from EU/EEA authorities.

Eligibility: what your device must meet

1. The device must be identical

The device supplied in Saudi Arabia must be identical to the one authorized by the reference authority. In practice, compare design, intended use and indications, components and materials, software version, manufacturer and manufacturing sites. A “Saudi variant” that differs from the authorized device may not qualify.

2. Saudi-specific requirements still apply

Even an identical device has to meet Saudi requirements that a foreign regulator does not check:

  • SBC 401 electrical requirements, where they apply to the device.
  • Labelling that meets SFDA rules.
  • IFU language: information for lay-use and self-test devices must be in Arabic and English. Professional IFU may be in English, with Arabic on request.
  • Storage requirements.

3. QMS evidence: MDSAP accepted

On the reliance pathway, SFDA accepts the Medical Device Single Audit Program (MDSAP) as quality management system evidence. ISO 13485:2016 remains the standard QMS requirement for MDMA applications generally.

What is excluded

MDS-G030 excludes these categories from reliance:

  • Self-declaration routes. For example, EU Class I devices that the manufacturer self-certified without a notified body. The point of reliance is that a regulator or its designated body actually assessed the device.
  • Predicate or 510(k)-type equivalence routes. Clearance based on substantial equivalence to a predicate device does not qualify.
  • Custom-made devices.
  • Devices subject to safety actions, such as recalls or field safety corrective actions.
  • Possibly software as a medical device (SaMD) and AI-based devices. How this exclusion applies is not fully clear. If your product is software or uses AI, confirm with SFDA before you plan around reliance.

Many US manufacturers rely on 510(k) clearance, so FDA status alone does not settle eligibility. Check exactly which authorization your device holds.

Reliance vs standard route: comparison

Reliance pathway (MDS-G030) Standard route
Legal basis MDS-G030 v1.0 (9 Sep 2026) plus MDS-REQ 1 MDS-REQ 1
Precondition Authorization by a reference authority for an identical device None
Depth of review Abridged; relies on reference decision Full technical documentation review
Stated review time 35–60 working days Not fixed in published guidance; confirm with SFDA
QMS evidence ISO 13485:2016 or MDSAP ISO 13485:2016
Saudi-specific requirements Required (SBC 401, labelling, IFU language, storage) Required
Excluded Self-declared, 510(k)/predicate, custom-made, safety-action devices; possibly SaMD/AI —
Authorized Representative Required for foreign manufacturers Required for foreign manufacturers
Result MDMA MDMA

Both routes produce the same MDMA, with the same validity rules (a maximum of 3 years) and the same post-market duties. For the full picture of the standard route, see our SFDA MDMA registration guide.

How to check if your device is eligible

Work through these questions in order:

  1. Which authority authorized the device, and through what route? Write down the exact authorization: certificate, licence or approval number and date.
  2. Was it a self-declaration or an equivalence route? If yes, plan for the standard route.
  3. Is the device under any safety action? Open recalls or FSCAs exclude it.
  4. Is it custom-made, software or AI-based? Custom-made is excluded. For SaMD/AI, confirm with SFDA first.
  5. Is the Saudi device identical? Compare the bill of materials, software version, labelling content and manufacturing sites.
  6. Can you meet the Saudi-specific requirements? Check SBC 401 for electrical devices, the Arabic/English labelling and IFU, and the storage conditions.
  7. Do you have a Saudi Authorized Representative? Foreign manufacturers apply through a licensed AR under MDS-REQ 9. See our Authorized Representative guide.

If you answer yes to the first and last steps and clear the exclusions, reliance is likely worth pursuing. Our approval checker gives a quick first view.

What reliance does not change

The reliance pathway shortens the technical review. It does not reduce your obligations before or after approval. Whichever route you take, these still apply:

  • Establishment licensing. The Saudi Authorized Representative must hold an establishment licence under MDS-REQ 9 v2.0. The AR needs a separate licence for each manufacturer, and only one AR is allowed per device type or group.
  • Declaration of Conformity on the SFDA template. The DoC follows Annex 14 of MDS-REQ 1. A foreign DoC format is not a substitute.
  • Validity and renewal. The MDMA is valid for a maximum of 3 years, or less if the underlying conformity certificates expire sooner. You can apply for renewal up to 90 days before expiry, and you cannot add new products at renewal.
  • Adverse event reporting. Under MDS-REQ 11, report incidents to the NCMDR within 2 days for a serious public-health threat, 10 days for death or serious injury, and 30 days for other events.
  • FSCA. Report a field safety corrective action within 2 days of the FSCA letter, then notify users and submit the plan within 5 days.
  • PSURs. Class C and D devices need an annual Periodic Safety Update Report.
  • UDI. Register the device in SFDA’s Saudi-DI database.

There is also a link between reliance and post-market events. Devices subject to safety actions are excluded from reliance. A recall or FSCA in another market can therefore change which route is open to you, so check the device’s safety history before you choose a route.

Who benefits most

Reliance suits manufacturers whose device:

  • already holds a full authorization from one of the six reference regulators;
  • is sold in the same configuration worldwide;
  • has bilingual labelling and electrical conformity work that can be done early.

Manufacturers whose main approval is a 510(k) clearance or an EU Class I self-declaration should plan for the standard route.

Practical notes for 2026

  • Portal naming. MDS-G030 refers to the “SFDA E-Services portal”. MDMA and licensing have run on SFDA’s GHAD system, and this may be a rename. Confirm the current entry point before you file.
  • Fees. Confirm reliance-route fees with SFDA rather than relying on figures circulating online.
  • Labelling is the usual bottleneck. Most of the work on a reliance file is Saudi-specific: bilingual labelling, IFU language and electrical conformity. Start these early.
  • Bundling of up to 50 items per application is available under MDS-REQ 1 Annex 15 and MDS-G028. If you plan to bundle under reliance, confirm with SFDA how the eligibility conditions apply to each item.

How Mutamad helps

Mutamad screens devices for reliance eligibility, prepares the Saudi-specific parts of the dossier and handles submission with your Authorized Representative. Services start from 999 SAR. See our medical device registration service or contact us with your reference authorization details.

This guide is general information reflecting our understanding of the requirements on the review date above. Regulations change; confirm current requirements on the official SFDA website or contact us about your specific case.

Frequently asked questions

What is the SFDA reliance pathway for medical devices?

+

It is a route under guidance MDS-G030 (v1.0, 9 September 2026) in which SFDA relies on a prior authorization from a recognised reference authority instead of carrying out a full technical documentation review. The device must be identical and meet Saudi-specific requirements.

Which regulators does SFDA rely on for medical devices?

+

MDS-G030 lists TGA (Australia), Health Canada, EU/EEA national competent authorities, PMDA (Japan), MHRA (UK) and the US FDA as reference authorities.

Can a 510(k)-cleared device use the SFDA reliance pathway?

+

Not as a rule. MDS-G030 excludes predicate or 510(k)-type equivalence routes, so a device cleared through 510(k) will normally need the standard route. Check the guidance or ask SFDA if your device has a different FDA authorization.

How long does SFDA reliance review take?

+

SFDA states a review time of 35 to 60 working days for reliance applications. The clock depends on a complete dossier, so missing Saudi-specific items such as Arabic labelling or SBC 401 evidence can add time.

Is MDSAP accepted by SFDA?

+

Yes, on the reliance pathway MDSAP is accepted as quality management system evidence. ISO 13485:2016 certification remains the general QMS requirement for MDMA applications.

Do software and AI medical devices qualify for reliance?

+

Possibly not. Software as a medical device and AI-based devices may be excluded from the reliance pathway, and the position should be confirmed with SFDA before planning a submission.