Medical devices
SFDA Medical Device Registration (MDMA) Saudi Arabia: 2026 Guide
Every medical device needs an SFDA Medical Device Marketing Authorization (MDMA) before supply in Saudi Arabia. Classes, documents, routes and steps for 2026.
Last reviewed: 9 min read By the Mutamad regulatory team
Every medical device sold in Saudi Arabia needs a Medical Device Marketing Authorization (MDMA) from the Saudi Food and Drug Authority (SFDA) before it can be supplied, and that includes Class A devices. Foreign manufacturers apply through a licensed Saudi Authorized Representative, submitting a technical file, ISO 13485:2016 evidence and an SFDA-format Declaration of Conformity. The MDMA is valid for up to 3 years. Since September 2026, devices already authorized in certain reference countries may be able to use a faster reliance pathway.
This guide covers how the MDMA works in 2026: classification, documentation, labelling, the Authorized Representative, the online system, validity and post-market obligations.
What is an MDMA?
The MDMA is the product-level marketing authorization SFDA issues for a medical device or in vitro diagnostic (IVD). The requirement comes from MDS-REQ 1 (Requirements for Medical Devices Marketing Authorization, v6, 19 Dec 2021), which says every device needs an MDMA before it is placed on the Saudi market. You can read the official text on the SFDA website.
Two other terms come up often:
- MDNR (Medical Device National Registry): the establishment or manufacturer registry number. It still appears on the SFDA Declaration of Conformity template (Annex 14).
- “MDEL”: a legacy shorthand for the establishment licence. The current instrument is the establishment licence under MDS-REQ 9 v2.0 (15 Jan 2025).
In short: the establishment licence covers the company, and the MDMA covers the product. You need both.
Device classification: Classes A to D
SFDA uses four risk classes, from A (lowest) to D (highest). General medical devices and IVDs both use Classes A–D, but under separate classification rules set out in MDS-REQ 1 Annex 5, with further guidance in MDS-G008.
Three points matter most in practice:
- Class A is not exempt. Unlike some markets where low-risk devices only need a self-declaration, a Class A device in Saudi Arabia still needs an MDMA.
- Class drives scrutiny. Higher classes need more clinical evidence and carry heavier post-market duties, such as annual PSURs for Class C and D.
- IVDs follow their own rules. Do not classify an IVD using the general device rules. The technical file format is different too (Annex 4 instead of Annex 3).
Reduced documentation for low-risk devices
MDS-REQ 1 (§4) allows a reduced documentation set for low-risk devices. It does not apply to:
- IVDs
- sterile devices
- devices with a measuring function
- reusable surgical instruments
- novel devices
Eligible devices submit only the device description, labelling, the Essential Principles (EP) checklist, evidence of EP conformity, the risk management file, and the post-market surveillance plan and report.
Two routes: standard review or reliance
| Standard route | Reliance pathway (MDS-G030) | |
|---|---|---|
| Basis | Full technical documentation review by SFDA | SFDA relies on a prior authorization by a reference authority |
| Who can use it | Any device | Identical device authorized by TGA, Health Canada, EU/EEA authority, PMDA, MHRA or US FDA |
| Excluded | — | Self-declaration routes, 510(k)/predicate-type equivalence routes, custom-made devices, devices with safety actions; possibly SaMD/AI |
| QMS evidence | ISO 13485:2016 | ISO 13485:2016; MDSAP accepted |
| Review time | Not fixed in published guidance; confirm with SFDA | 35–60 working days |
| Saudi-specific checks | Yes | Yes: SBC 401 electrical, labelling, IFU language, storage |
The standard route is a full technical documentation review, sometimes called a “TFA”. It works for any device, including those with no foreign approval.
The reliance pathway is new. MDS-G030 v1.0 was published on 9 September 2026. It lets SFDA rely on a previous authorization from a recognised reference authority, provided the device is identical and meets Saudi-specific requirements. Read our full reliance pathway guide for eligibility details.
Documents you need for an MDMA
For the standard route, expect to prepare:
- Technical documentation following Annex 3 of MDS-REQ 1 for general devices, or Annex 4 for IVDs.
- Quality management system evidence: ISO 13485:2016 certification.
- Declaration of Conformity (DoC) on the SFDA template (Annex 14). A manufacturer’s generic EU or other DoC is not a substitute.
- Clinical evaluation, including post-market clinical follow-up where relevant (Annex 6), or for IVDs, a performance evaluation (Annex 7).
- Labelling and instructions for use (IFU) as they will be supplied in Saudi Arabia.
- Risk management file and post-market surveillance plan.
Bundling: up to 50 items per application
Related products can be grouped. SFDA allows up to 50 items per bundled application (MDS-REQ 1 Annex 15; guidance MDS-G028). Bundling well cuts administration, so plan your groupings before you start the technical file.
Arabic labelling requirements
SFDA’s language rules depend on who uses the device:
- Lay-use and self-test devices: the information supplied with the device must be in Arabic and English.
- Advertising to the public: must be bilingual.
- Professional-use devices: the IFU may be in English only, with an Arabic version available on request.
Mismatches between the submitted labelling and the product as shipped are a common cause of queries. If you need a second pair of eyes on bilingual artwork, see our Arabic label review service.
The Authorized Representative and establishment licence
A foreign manufacturer cannot apply for an MDMA directly. Under MDS-REQ 9, it must apply through a licensed Saudi Authorized Representative (AR). The AR must:
- be a Saudi legal entity;
- hold ISO 13485 evidence or an SFDA inspection report;
- have a qualified contact person;
- hold a separate licence for each manufacturer it represents;
- have a written agreement with the manufacturer containing the mandatory terms.
Only one AR per device type or group is allowed. The AR’s duties continue after approval: annual confirmation of information, post-market surveillance cooperation, reporting foreign incidents, notifying NCMDR of shortages and ensuring after-sales service and spare parts. Its liability continues until a replacement AR is appointed.
The AR licence fee published by SFDA is SAR 2,600 per year, for a term of 1–10 years. Our Authorized Representative guide covers the role in detail.
Where you apply: GHAD
Establishment licensing and MDMA applications run on SFDA’s unified electronic system, GHAD. The September 2026 reliance guidance (MDS-G030) refers to the “SFDA E-Services portal”. This may simply be a new name for the same system. Check the current entry point with SFDA or your AR before you file.
Validity, renewal and fees
- Validity: an MDMA lasts a maximum of 3 years. It will be shorter if the underlying conformity certificates expire sooner.
- Renewal: apply up to 90 days before expiry. You cannot add new products in a renewal; new products need a new application.
- Fees: SFDA fees depend on the device class and the route. A 2025 SFDA fee sheet circulates online, but its status is unclear (it may relate to an optional expedited service). Confirm current fees directly with SFDA before budgeting.
Post-market obligations
An MDMA starts your compliance obligations; it doesn’t end them.
Adverse event reporting to NCMDR
Under MDS-REQ 11 v2 (23 Mar 2023), incidents go to the National Center for Medical Devices Reporting (NCMDR):
| Event | Reporting deadline |
|---|---|
| Serious public-health threat | 2 days |
| Death or serious injury | 10 days |
| Other reportable events | 30 days |
| Response to NCMDR queries | 5 days |
Field safety corrective actions (FSCA)
Report an FSCA to SFDA within 2 days of the FSCA letter. Then notify users and submit the action plan within 5 days.
PSURs
Class C and D devices need an annual Periodic Safety Update Report (PSUR).
UDI and Saudi-DI
Devices must be registered in SFDA’s UDI database, Saudi-DI (udi.sfda.gov.sa). SFDA’s Announcement of 01/08/2021 set these compliance dates:
- Classes B, C and D: 1 September 2022
- Class A: 1 September 2023
Some sources cite later dates (2023 for B–D and 2024 for A). We found no newer SFDA circular confirming this. In any case, both sets of dates have now passed, so assume UDI applies to your device and confirm the details with SFDA.
MDMA at a glance
| Topic | Requirement |
|---|---|
| Legal basis | MDS-REQ 1 (v6, Dec 2021) |
| Who needs it | Every medical device and IVD, Classes A–D |
| Applicant | Saudi Authorized Representative (for foreign manufacturers) |
| Company licence | Establishment licence under MDS-REQ 9 v2.0 |
| QMS | ISO 13485:2016 (MDSAP accepted on reliance route) |
| DoC | SFDA template, Annex 14 |
| Technical file | Annex 3 (devices) / Annex 4 (IVDs) |
| Bundling | Up to 50 items per application |
| Validity | Max 3 years; renew up to 90 days before expiry |
| Arabic | Required for lay/self-test info; professional IFU may be English |
| Post-market | NCMDR 2/10/30 days, FSCA, PSUR (C/D), Saudi-DI |
How Mutamad helps
Mutamad prepares and submits MDMA applications, reviews Arabic labelling and supports manufacturers with Authorized Representative and establishment licensing arrangements. Services start from 999 SAR. See our medical device registration service, check your product with the approval checker, or contact us to discuss your device.