Saudi Authorized Representative for Medical Devices (2026)
Foreign medical device makers need a licensed Saudi Authorized Representative under SFDA MDS-REQ 9. Conditions, duties, the SAR 2,600/year fee and choosing one.
Last reviewed: 9 min read By the Mutamad regulatory team
If you manufacture medical devices outside Saudi Arabia, you need a licensed Saudi Authorized Representative (AR) before you can sell there. SFDA’s establishment licensing requirements, MDS-REQ 9 v2.0 (15 January 2025), say a foreign manufacturer applies to SFDA through an AR. That AR must be a Saudi legal entity with quality-system evidence and a written agreement with you. It holds a licence specific to your company, which costs SAR 2,600 per year.
This guide explains who can act as your AR, what the role involves, what it costs, and how to choose between your distributor and an independent AR.
Why foreign manufacturers need a Saudi AR
SFDA regulates medical devices through two layers. First, every device needs a Medical Device Marketing Authorization (MDMA) before supply, including Class A devices (MDS-REQ 1). Second, the companies that place devices on the market need establishment licences under MDS-REQ 9 v2.0.
A manufacturer with no Saudi legal presence cannot hold that local accountability itself. MDS-REQ 9 §2 therefore requires the foreign manufacturer to apply through a licensed Saudi AR. In practice, the AR is the company that files your MDMA applications in SFDA’s system (GHAD), answers SFDA on your behalf, and carries legal responsibility in the Kingdom.
For the full registration process, see our guide to SFDA medical device registration (MDMA). If your device is already cleared by a recognised reference regulator, the SFDA reliance pathway can shorten review, but you still need an AR to file.
Who can be an Authorized Representative
MDS-REQ 9 sets six conditions. All of them must be met before the AR can be licensed for your company.
| Condition | What it means in practice |
|---|---|
| Saudi legal entity | The AR must be a legal entity established in Saudi Arabia. A company based abroad cannot fill the role. |
| ISO 13485 evidence or SFDA inspection report | The AR must show a quality management system, either through ISO 13485 evidence or an SFDA inspection report. |
| Qualified contact person | A named, qualified person who handles SFDA communication, vigilance and regulatory questions. |
| Separate licence per manufacturer | If the AR represents five manufacturers, it holds five AR licences. Your licence covers your company only. |
| One AR per device type or group | You cannot appoint two ARs for the same device type or group. Different groups can have different ARs. |
| Written agreement with mandatory terms | The appointment must be documented in a signed agreement that contains the terms MDS-REQ 9 requires. |
The “one AR per device type or group” rule matters strategically. Whoever you appoint becomes the only route to SFDA for that product line until you replace them.
AR licence fee and term
SFDA charges SAR 2,600 per year for the AR licence, and the licence can run for a term of 1 to 10 years. Because a separate licence is needed for each manufacturer, this fee applies to your company specifically, not to the AR’s business as a whole.
Keep three points in mind:
- The SAR 2,600 is SFDA’s government fee. The AR’s own service fee is a separate commercial matter.
- A longer licence term reduces renewal administration, but agree in the contract what happens to prepaid years if you end the relationship early.
- MDMA fees are separate from the AR licence. See the MDMA guide for the current fee position, which should be confirmed with SFDA.
What the AR is responsible for
The AR is not a mailbox. MDS-REQ 9 gives it ongoing regulatory duties for as long as it is appointed.
Annual confirmation of information
The AR must confirm your registered information with SFDA every year. Out-of-date manufacturer details, addresses or contacts can cause problems at renewal or during an inspection, so give your AR a clear channel for updates.
Representing you before SFDA
The AR answers SFDA queries, receives regulatory correspondence and acts for you in applications. Its response times directly affect your review timelines.
Post-market surveillance cooperation
The AR cooperates with you on post-market surveillance (PMS). Adverse events in Saudi Arabia are reported to SFDA’s National Center for Medical Devices Reporting (NCMDR) under MDS-REQ 11 v2. The deadlines are short: 2 days for a serious public-health threat, 10 days for death or serious injury, and 30 days for other events. NCMDR queries must be answered within 5 days. Field safety corrective actions must be reported within 2 days of the FSCA letter.
Reporting foreign incidents
The AR must also report relevant incidents that occur outside Saudi Arabia. Your global vigilance team needs to feed the AR promptly, not just your Saudi distributor.
Shortage notification
The AR must notify NCMDR of supply shortages. If you plan to discontinue a product or expect supply problems, tell your AR early.
After-sales service and spare parts
The AR must make sure after-sales service and spare parts are available. For capital equipment this is often handled by a distributor or service partner, but the AR remains accountable for it being in place.
Liability until replacement
The AR’s responsibility continues until a replacement AR is appointed. This protects patients from gaps in representation. It also means an AR may be cautious about letting go without a clean handover.
Distributor-as-AR vs independent AR
Many manufacturers appoint their main Saudi distributor as AR because it is convenient. Others keep regulatory representation separate from sales. Both are allowed if the AR meets the MDS-REQ 9 conditions.
| Factor | Distributor as AR | Independent AR |
|---|---|---|
| Setup speed | Often fast if the distributor already holds ISO 13485 evidence or an SFDA inspection report | Fast if the independent AR is already licensed for other manufacturers |
| Commercial leverage | Distributor controls both sales and your SFDA representation | Regulatory role is separate from sales; you keep negotiating leverage |
| Changing distributors | Harder: changing distributor also means changing AR for that device group | Easier: add or switch distributors without touching your AR or MDMAs |
| Multiple distributors | Awkward, since one AR per device type or group | Natural fit: one neutral AR, several distributors |
| Vigilance and PMS | Close to the market and customers | Needs good data flows from distributors to the AR |
| After-sales and spare parts | Often handled in-house | Must be arranged with distributors or service partners and documented |
| Cost | Sometimes bundled into the distribution margin | Separate service fee, but transparent |
A common middle path is to start with an independent AR while you test distributors, then decide later whether to consolidate.
How to change your Authorized Representative
Changing AR is possible, but plan it as a project rather than a termination letter.
- Appoint and license the new AR first. The new AR needs its own licence for your company before it can take over.
- Agree the handover. Transfer technical documentation, open vigilance cases, complaint files and PMS data.
- Update SFDA records. The MDMAs and your manufacturer information need to point to the new AR. Confirm the exact steps with SFDA, as the procedure can change.
- End the old agreement. The outgoing AR stays responsible until the replacement is appointed, so time termination to avoid any gap.
Because only one AR can cover a device type or group, you cannot simply run two in parallel for the same products during the transition.
What to put in the AR agreement
MDS-REQ 9 requires a written agreement with mandatory terms. Check the current text of MDS-REQ 9 v2.0 for the exact wording SFDA expects. In practice, a robust agreement should also cover:
- Scope: which device types or groups and which MDMAs the AR covers.
- Term: aligned with the AR licence term (1 to 10 years), with renewal terms.
- Regulatory duties: annual information confirmation, SFDA correspondence and response times.
- Vigilance: how incidents in Saudi Arabia and abroad flow between you and the AR, aligned with the NCMDR deadlines.
- Shortages: advance notice of discontinuations or supply problems.
- After-sales and spare parts: who provides service and how the AR verifies it.
- Document access: the AR’s access to technical documentation and your right to retrieve it.
- Termination and handover: notice periods, cooperation on transfer to a new AR, and return of documents and system access.
- Fees: the SFDA licence fee and the AR’s service fee, stated separately.
How Mutamad helps
Mutamad can help you set up and manage a Saudi Authorized Representative arrangement, review your AR agreement, and coordinate MDMA filings through our medical device registration service. Services start from 999 SAR; see pricing or contact us to discuss your device portfolio.
This guide is general information reflecting our understanding of the requirements on the review date above. Regulations change; confirm current requirements on the official SFDA website or contact us about your specific case.